The HI96710 portable photometer is for the measurement of pH, free
chlorine, and total chlorine. Hanna's portable photometers feature an
advanced optical system; the combination of a special tungsten lamp, a
narrow band interference filter, and silicon photodetector ensure
accurate photometric readings every time. The Hanna exclusive CAL Check
™
feature utilizes ready-made, NIST traceable standards to verify both
meter validation and calibration. The exclusive cuvette locking system
ensures that the cuvette is inserted into the measurement cell in the
same position every time to maintain a consistent path length.
Features at-a-glance
CAL Check™ - Allows for performance verification and calibration of the meter using a CAL Check NIST traceable secondary standard.
GLP - Records the date of the last user calibration performed.
Built in timer - Ensures the appropriate
reaction time is used for the chemical reaction. At the end of the timer
the meter will automatically take the reading. This feature ensures
consistency across multiple users.
Indexing indent on meter - The meter has
an indent that allows for a lock and key fit with the cuvette cap. This
ensures that the cuvette is indexed consistently (same position) in
order to maintain the same path length for accurate results.
Cooling Lamp Indicator - It is necessary to maintain
a consistent temperature of the optical components in order to maintain
a narrow wavelength band of light. This photometer has a cooling lamp
indicator that is displayed for a short period of time before each
measurement in order to allow the components to cool and obtain the
highest accuracy possible.
Error messages - Messages on display alerting to problems including no cap, over range and under range readings, and light source error.
Auto-shut off - Automatic shut off after 10 minutes
of non-use when the meter is in measurement mode. Prevents wastage of
batteries in the event the meter is accidentally not turned off.
Battery status indicator
The instrument has a battery level indicator to display reaming life as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinks when the capacity is less than 10 %.
Units of measure - Appropriate unit of measure is displayed along with reading.
Three critical parameters that can be tested to ensure good water
quality are pH, free chlorine, and total chlorine. Chlorine is one of
the most commonly used disinfectants for drinking water, wastewater, and
water used for pools and spas. It can be added to in various forms
including calcium hypochlorite, sodium hypochlorite, or in some
instances, chlorine gas. When added to water, chlorine creates
hypochlorous acid (HOCl) which dissociates into hypochlorite ion (OCl
-).
HOCl ↔ H+ + OCl-
hypochlorous acid ↔ hydrogen ion + hypochlorite ion
HOCl is the form of chlorine that acts as a stronger disinfectant as compared to OCl
-.
To ensure the added chlorine is effective at sanitizing, the pH of the
water must be taken into account. Around pH 7.5, HOCl and OCl
-
are present in relatively equal amounts. Below pH 7.5, the equilibrium
shifts to favor HOCl; above pH 7.5, the equilibrium shifts to favor
OCl-. Depending on the application, addition of chlorine is effective
when added to water with a neutral or slightly acidic pH value.
When chlorine is first added to water, it is available as free
chlorine. The measurement of free chlorine signifies the amount
available for disinfection. Once chlorine begins to sanitize bacteria
and pathogens present in the water, it becomes combined chlorine;
combined chlorine is no longer available to act as a disinfectant. The
measurement of total chlorine signifies the amount of free chlorine and
combined chlorine. With both free and total chlorine measurements, a
drinking water operator or pool owner can determine if there is enough
chlorine available for disinfection.
The HI96710 offers multiple methods within one meter. pH
determination uses the phenol red method while free and total chlorine
determination uses an adaptation of the EPA method 330.5 and Standard
Method 4500-Cl G. The reagents are in powder and liquid form depending
on the parameter being tested.
When the appropriate reagent is added to the sample, it will display a
color change; the greater the concentration, the deeper the color. The
associated color change is then colorimetrically analyzed according to
the Beer-Lambert Law. This principle states that light is absorbed by a
complementary color, and the emitted radiation is dependent upon
concentration. For determination of pH and free and total chlorine, a
narrow band interference filter at 525 nm (green) allows only green
light to be detected by the silicon photodetector and omits all other
visible light emitted from the tungsten lamp. As the change in color of
the reacted sample increases, absorbance of the specific wavelength of
light also increases, while transmittance decreases.
The HI96710 is available as a kit (HI96710C) which includes:
- Portable Photometer
- Sample cuvettes (2)
- Caps
- CAL Check™ standards with certificate
- Cuvette cleaning cloth
- Scissors
- Rugged carrying case
- Carrying case has thermoformed insert for meter and accessories
*Reagents ordered separately
CAL Check™ standards with certificate
The CAL Check™ standards are used for calibration and performance verification of photometers with the CAL Check™ function. Individual sets of CAL Check™standards are available for each parameter found in the HI96710:
- pH: HI96710-11
- Free Chlorine: HI96701-11
- Total Chlorine: HI96711-11
Supplied with Certificate of Analysis
- Lot number
- Expiration date
- Standard value @ 25 °C
- Reference meter NIST traceable
Provided storage containers
- Light tight
- Protects from accidental breakage